Область применения

1.1 This test method covers the determination of the concentration of technetium-99 in urine using inductively coupled plasma-mass spectrometry (ICP-MS). This test method can be used to support uranium enrichment and reclamation facility bioassay programs.

1.2 The minimum detectable concentration for this test method, using a quadrupole ICP-MS, is approximately 1.0 ng/L (0.62 Bq/L).

Ключевые слова:
bioassay; inductively coupled plasma-mass spectrometry; mass; technetium; urine