(812) 309-78-59
(495) 223-46-76
ASTM D5096-02
Standard Test Method for Determining the Performance of a Cup Anemometer or Propeller Anemometer
5 стр.
Заменен
Электронный (pdf)Печатное издание
96.72 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
07.060 Geology. Meteorology. Hydrology / Геология. Метеорология. Гидрология
Сборник (ASTM):
11.07 Air Quality / Атмосферный анализ
Тематика:
Atmospheric Analysis
Описание
Значение и использование

This test method will provide a standard for comparison of rotating type anemometers, specifically cup anemometers and propeller anemometers, of different types. Specifications by regulatory agencies (4-7) and industrial societies have specified performance values. This standard provides an unambiguous method for measuring Starting Threshold, Distance Constant, Transfer Function, and Off-Axis Response.

Область применения

1.1 This test method covers the determination of the Starting Threshold, Distance Constant, Transfer Function, and Off-Axis Response of a cup anemometer or propeller anemometer from direct measurement in a wind tunnel.

1.2 This test method provides for a measurement of cup anemometer or propeller anemometer performance in the environment of wind tunnel flow. Transference of values determined by these methods to atmospheric flow must be done with an understanding that there is a difference between the two flow systems.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
anemometer; distance constant; off-axis response; threshold; wind speed