ASTM E1270-88(2014)
Standard Test Method for Equal Arm Balances (Withdrawn 2021)
6 стр.
Отменен
Электронный (pdf)Печатное издание
96.72 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
17.100 Measurement. Including measuring and weighing instruments / Измерения силы, веса и давления. Включая измерительные приборы
Сборник (ASTM):
14.04 Laboratory Apparatus; Degradation of Materials; SI; Oxygen Fire Safety / Лабораторная аппаратура; Деградация материалов; Международная система единиц СИ; Кислород и пожарная безопасность
Тематика:
Laboratory Testing
Описание
Причина отменыThis test method can be used for testing equal-arm balances of any capacity and sensitivity. The testing procedure should enable the user to characterize his instrument sufficiently to determine whether or not it is suitable for the purpose for which it is to be used.

Formerly under the jurisdiction of Committee E41 on Laboratory Apparatus, this test method was withdrawn in July 2021 and replaced by Practice E898 on the Calibration of Non-Automatic Weighing Instruments.1

Значение и использование

5.1 This test method should enable the user of the balance to interpret data determined thereon in terms of accuracy and precision. It should be helpful in using a particular instrument to best advantage. Weaknesses as well as strengths should become apparent. It is not the intention of this test method to compare similar instruments of different manufacture but rather to assist in choosing an instrument which will meet the needs of the user.

Область применения

1.1 This test method can be used for testing equal-arm balances of any capacity and sensitivity. The testing procedure should enable the user to characterize his instrument sufficiently to determine whether or not it is suitable for the purpose for which it is to be used.

1.2 The characteristics to be examined include:

1.2.1 Sensitivity at all loads,

1.2.2 Lever arm ratio,

1.2.3 Damping ratio (for instruments without accessory dampers),

1.2.4 Period of oscillation,

1.2.5 Precision, and

1.2.6 Linearity and calibration of accessory devices that provide on-scale indication of weight.

1.3 This standard does not purport to address all of the safety concerns associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.