(812) 309-78-59
(495) 223-46-76
ASTM E1772-95(2001)
Standard Test Method for Particle Size Distribution of Chromatography Media by Electric Sensing Zone Technique (Withdrawn 2010)
3 стр.
Отменен
Электронный (pdf)Печатное издание
84.24 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
19.120 Particle. Including test sieves and porosimetry / Ситовый анализ. Просеивание. Включая контрольные сита и измерение пористости
Сборник (ASTM):
11.06 Environmental; Biological Effects and Environmental Fate; Industrial Biotechnology / Окружающая среда; Биологические эффекты и трансформация в окружающей среде; Промышленные биотехнологии
Тематика:
Biotechnology
Описание
Причина отмены

This test method is valuable for the measurement of particle size and covers determination of the particle size distribution of chromatography media in the overall size range of approximately 1 to 450 μm using the electric sensing zone (ESZ) apparatus. This instrument uses an electric current path of small dimensions that is modulated by individual particle passage through an aperture and produces individual pulses of amplitude proportional to the particle volume.

Formerly under the jurisdiction of Committee E48 on Biotechnology, this test method was withdrawn in February 2010 in accordance with section 10.5.3.1 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.

Значение и использование

It is important to recognize that the results obtained by this test method or any other test method for particle size determination using different physical principles may disagree. The results are strongly influenced by physical principles used by each method of particle size analysis. The results of any particle sizing method should be used only in a relative sense and should not be regarded as absolute when comparing results obtained by other methods.

Both sellers and purchasers of chromatography media will find the test method useful if mutually acceptable to all parties concerned to determine the particle size distributions for materials specifications, manufacturing control, and development and research.

Область применения

1.1 This test method is valuable for the measurement of particle size and covers determination of the particle size distribution of chromatography media in the overall size range of approximately 1 to 450 μm using the electric sensing zone (ESZ) apparatus. This instrument uses an electric current path of small dimensions that is modulated by individual particle passage through an aperture and produces individual pulses of amplitude proportional to the particle volume (1).²

1.2 The values stated in SI units are to be regarded as the standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
Chromatography media; Electric sensing zone; particle size