Описание
Значение и использование

This test method provides for accurate and reproducible enumeration of particles and fibers released from a wiper immersed in a cleaning solution with moderate mechanical stress applied. When performed correctly, this counting test method is sensitive enough to quantify very low levels of total particle and fiber burden. The results are accurate and not influenced by artifact or particle size limitations. A further advantage to this technique is that it allows for morphological as well as X-ray analysis of individual particles.

Область применения

1.1 This test method covers testing all wipers used in cleanrooms and other controlled environments for characteristics related to particulate cleanliness.

1.2 This test method includes the use of computer-based image analysis and counting hardware and software for the counting of densely particle-laden filters (see 7.7-7.9). While the use of this equipment is not absolutely necessary, it is strongly recommended to enhance the accuracy, speed, and consistency of counting.

1.3 The values stated in SI units are to be regarded as the standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
cleanroom wipers; contamination control; membrane filter; optical microscopy; particle counting and sizing; particles and fibers; scanning electron microscopy; Cleanrooms (materials/applications); Fiber analysis; Optical electron microscopy; Particle counters/counting; Scanning electron microscope (SEM); Size-diferentiated counting