(812) 309-78-59
(495) 223-46-76
ASTM F1925-09
Standard Specification for Semi-Crystalline Poly(lactide) Polymer and Copolymer Resins for Surgical Implants
9 стр.
Заменен
Электронный (pdf)Печатное издание
96.72 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
11.040.40 Implants. Including pacemakers / Имплантаты для хирургии, протезирования и ортоптики. Включая кардиостимуляторы83.080.20 Thermoplastic materials / Термопластические материалы
Сборник (ASTM):
13.01 Medical and Surgical Materials and Devices (I): E667 – F2477 / Медицинские и хирургические материалы и приборы (I): E667 - F2477
Тематика:
Medical Devices & Implants
Описание
Реферат

This specification covers virgin poly(L-lactic acid) resin (PLLA resin) intended for use in surgical implants. This specification does not cover stereoisomeric compositions based on various D, L, or DL copolymer ratios. This specification addresses material characteristics of virgin poly(L-lactic acid) resin and does not apply to packaged and sterilized finished implants fabricated from this material. The virgin polymer shall be a homopolymer of L-lactide with the prescribed density. The molecular mass of the virgin polymer shall be indicated by relative solution viscosity (in chloroform). In addition, the weight average molecular mass and molecular mass distributions may be determined by gel permeation chromatography The virgin polymer shall be identified as a polylactide by infrared or 1H-NMR spectroscopy. Typical infrared transmission and 1H-NMR spectra are shown. The virgin polymer shall have a specific optical rotation (in dichloromethane) and residual monomer content within the prescribe values, and shall conform to the chemical and physical property requirements specified for: residual solvent, residual water, residual tin, heavy metals, and sulfated ash. The following test methods shall be used: (1) Karl-Fischer titration and (2) atomic absorption-emission (AA) spectroscopy or inductively coupled plasma (ICP) spectroscopy. Considerations for biocompatibility of the material from a human implant perspective is also given.

Область применения

1.1 This specification covers virgin semi-crystalline poly(l-lactide) or poly(d-lactide) homopolymer resins intended for use in surgical implants. This specification also covers semi-crystalline resins of l-lactide copolymerized with other bioabsorbable monomers including, but not limited to, glycolide, d-lactide, and dl-lactide. The poly(l-lactide) or poly(d-lactide) based homopolymers and copolymers covered by this specification possess lactide segments of sufficient length to allow potential for their crystallization upon annealing.

1.2 Since poly(glycolide) is commonly abbreviated as PGA for poly(glycolic acid) and poly(lactide) is commonly abbreviated as PLA for poly(lactic acid), these polymers are commonly referred to as PGA, PLA, and PLA:PGA resins for the hydrolytic byproducts to which they respectively degrade. PLA is a term that carries no stereoisomeric specificity and therefore encompasses both the amorphous atactic/syndiotactic dl-lactide based polymers and copolymers as well as the isotactic d-PLA and l-PLA moieties, each of which carries potential for crystallization. Inclusion of stereoisomeric specificity within the lactic acid based acronyms results in the following: poly(l-lactide) as PlLA for poly(l-lactic acid), poly(d-lactide) as PdLA for poly(d-lactic acid), and poly(dl-lactide) as PdlLA for poly(dl-lactic acid).

1.3 This specification is applicable to lactide-based polymers or copolymers that possess isotactic polymeric segments sufficient in size to carry potential for lactide-based crystallization. Such polymers typically possess nominal mole fractions that equal or exceed 50 % l-lactide. This specification is particularly applicable to isotactic-lactide based block copolymers or to polymers or copolymers synthesized from combinations of d-lactide and l-lactide that differ by more than 1.5 total mole percent (1.5 % of total moles). This specification is not applicable to lactide-co-glycolide copolymers with glycolide mole fractions greater than or equal to 70 % (65.3 % in mass fraction), which are covered by Specification F 2313. This specification is not applicable to amorphous polymers or copolymers synthesized from combinations of d-lactide and l-lactide that differ by less than 1.5 total mole percent (1.5 % of total moles) as covered by Specification F 2579.

1.4 This specification covers virgin semi-crystalline poly(lactide)-based resins able to be fully solvated at 30°C by either methylene chloride (dichloromethane) or chloroform (trichloromethane). This specification is not applicable to lactide:glycolide copolymers that possess glycolide segments sufficient in size to deliver potential for glycolide-based crystallization, thereby requiring fluorinated solvents for complete dissolution under room temperature conditions (see Specification F 2313).

1.5 Within this specification, semi-crystallinity within the resin is defined by the presence of a DSC (differential scanning calorimetry) crystalline endotherm after annealing above the glass transition temperature. While other copolymeric segments may also crystallize upon annealing (for example, glycolide), specific characterization of crystalline structures other than those formed by lactide are outside the scope of this specification.

1.6 This specification addresses material characteristics of the virgin semi-crystalline poly(lactide) based resins intended for use in surgical implants and does not apply to packaged and sterilized finished implants fabricated from these materials.

1.7 As with any material, some characteristics may be altered by processing techniques (such as molding, extrusion, machining, assembly, sterilization, and so forth) required for the production of a specific part or device. Therefore, properties of fabricated forms of this resin should be evaluated independently using appropriate test methods to assure safety and efficacy.

1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
PGA:PLA; PLA; PLA:PGA; PLGA; PlLA; polyglycolic:lactic acid; poly(glycolide:lactide); poly(lactic acid); polylactic:glycolic acid; polylactide; poly(lactide); poly(lactide:glycolide); poly(l-lactic acid); Plastic surgical devices/applications--specifications; PLLA (poly(L-lactic acid)) resin; Polymers (surgical applications)--specifications; Surgical implant resins--specifications; Surgical implants--specifications