(812) 309-78-59
(495) 223-46-76
ASTM F2082-06
Standard Test Method for Determination of Transformation Temperature of Nickel-Titanium Shape Memory Alloys by Bend and Free Recovery (Withdrawn 2015)
7 стр.
Заменен
Электронный (pdf)Печатное издание
96.72 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
77.120.50 Titanium and titanium alloys / Титан и титановые сплавы
Сборник (ASTM):
13.01 Medical and Surgical Materials and Devices (I): E667 – F2477 / Медицинские и хирургические материалы и приборы (I): E667 - F2477
Тематика:
Medical Devices & Implants
Описание
Причина отменыThis test method described a procedure for determining the martensite-to-austenite transformation temperatures of either fully-annealed or heat-treated nickel titanium alloys by measuring the deformation recovered during the thermal transformation.

Formerly under the jurisdiction of Committee F04 on Medical and Surgical Materials and Devices, this practice was withdrawn in January 2015 in accordance with section 10.6.3 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.

Значение и использование

This test method provides a rapid, economical method for determination of transformation temperatures.

Measurement of the specimen motion closely parallels many shape memory applications and provides a result that is applicable to the function of the material.

This test method uses wire, tube, or strip samples; thus, it is able to provide an assessment of the product in its semifinished form.

This test method may be used on annealed samples to determine the transformation temperatures and assure the alloy formulation, since chemical analysis is not precise enough to determine adequately the nickel-to-titanium ratio of shape memory alloys.

Transformation temperatures derived from this test method may differ from those derived from other methods as a result of effects of strain and load on the transformation temperature.

The test method is applicable to shape memory alloys with Af temperatures in the range of approximately 25 to +90°C.

Область применения

1.1 This test method describes a procedure for determining the martensite-to-austenite transformation temperatures of either fully-annealed or heat-treated nickel titanium alloys by measuring the deformation recovered during the thermal transformation.

1.2 The values in SI units are to be regarded as the standard. The values given in inch-pound units are provided for information only.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
free recovery; nickel titanium; nitinol; shape memory; transformation temperature