Описание
Значение и использование

These test methods can be used to determine the effects of head and cone materials, design variables, manufacturing, and other conditions on the static and cyclic load-carrying ability of modular femoral heads mounted on the cones of femoral stem prostheses.

These test methods may use actual femoral prostheses or neck-cone models of simplified geometry with the same geometrical and material characteristics as in the implants. In either case, the matching metallic cone region of the test specimen selected shall be of the same material, tolerances, and finishing as the final femoral stem prosthesis.

The static test data may yield valuable information about the relative strengths and merits of different head and cone designs for particular applications. Due to the high forces anticipated for this type of destructive test (>40 kN), the boundary conditions and load levels far exceed possible in vivo loading parameters and therefore may not necessarily be applicable as a quantitative indicator of expected in vivo device performance.

In the fatigue test methods, it is recognized that actual loading in vivo is quite varied, and that no one set of experimental conditions can encompass all possible variations. Thus, the test methods included here represent a simplified model for the purposes of comparisons between designs and materials. These test methods are intended to be performed in air.

The test data may yield valuable information about the relative strengths of different head and cone designs.

Область применения

1.1 These test methods cover the evaluation of the static and cyclic fatigue strength of ceramic modular femoral heads, mounted on a cone as used on the femoral stem of the total hip arthroplasty.

1.2 These test methods were primarily developed for evaluation of ceramic (Specifications F 603 and F 1873) head designs on metal cones but may have application to other materials.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
bore; ceramic; cone; fatigue; modular head; static; strength; Ceramic head designs; Cyclic fatigue testing; Fatigue--surgical implant materials; Femoral surgical components; Metal cones/conical fittings; Modular prostheses; Static fatigue testing