(812) 309-78-59
(495) 223-46-76
ASTM D5996-96(2000)
Standard Test Method for Measuring Anionic Contaminants in High-Purity Water by On-Line Ion Chromatography
5 стр.
Заменен
Печатное изданиеЭлектронный (pdf)
93,60 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
71.040.50 Physicochemical. Including spectrophotometric and chromatographic analysis / Физико-химические методы анализа. Включая спектрофотометрический и хроматографический анализы
Сборник (ASTM):
11.01 Water (I) / Вода (I)
Тематика:
Water Testing
Описание
Область применения

1.1 This test method covers on-line analysis of high-purity water by the ion chromatography technique. This test method is applicable for measuring various anionic contaminants in high-purity water, typically in the range of 0.02 to 100 µg/L. This test method is used to determine the concentration of acetate, formate, chloride, fluoride, phosphate, nitrate, and sulfate in a continuously flowing sample. The range of the test method is only as good as the reagent water available for preparing standards. At extremely low concentrations, <1.0 µg/L, preparing standards is difficult, and extra care must be taken in their preparation. The sample may have to be conditioned from higher pressures and temperatures to conditions that are suitable for use by on-line instruments. The range of the test method is only as good as the reagent water available for standard preparation.

1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

Ключевые слова:
anions; high purity; ion; chromotography; on-line