(812) 309-78-59
(495) 223-46-76
ASTM E839-05
Standard Test Methods for Sheathed Thermocouples and Sheathed Thermocouple Material
14 стр.
Заменен
Электронный (pdf)Печатное издание
107.64 $ (включая НДС 20%)
Разработчик:
Зарубежные/ASTM
ICS:
17.200.20 Temperature-measuring. Including thermostats / Приборы для измерения температуры. Включая термостаты
Сборник (ASTM):
14.03 Sensory Evaluation; Temperature Measurement; Language Services and Products / Сенсорная оценка; Измерение температуры; Лингвистические услуги и продукты
Тематика:
Temperature Measurement
Описание
Значение и использование

This test method provides a description of test methods used in ASTM specifications that establish certain acceptable limits for characteristics of thermocouple assemblies and thermocouple cable.

The intended use of these test methods is to define the methods by which the characteristics shall be determined.

The usefulness and purpose of the included tests are given for the category of tests.

Область применения

1.1 These are test methods for sheathed thermocouple material and assemblies.

1.2 The tests are intended to ensure quality control and to evaluate the suitability of sheathed thermocouple material or assemblies for specific applications. Some alternative test methods to obtain the same information are given, since in a given situation an alternative test method may be more practical. Service conditions are widely variable and so it is unlikely that all the tests described will be appropriate for a given thermocouple application. A brief statement is made following each test description to indicate when it might be used.

1.3 The tests described herein include test methods to measure the following properties of sheathed thermocouple material and assemblies.

1.3.1 Insulation PropertiesCompaction-absorption method and tension method.

Thickness.

Resistance-room temperature and elevated temperature.

1.3.2 Sheath PropertiesIntegrity-water test (two test methods) and mass spectrometer.

Dimensions-length, diameter, and roundness.

Wall thickness.

Surface-gross visual, finish, defect detection by dye penetrant, and cold-lap detection by tension test.

Metallurgical structure.

Ductility-bend test and tension test.

1.3.3 Thermoelement PropertiesCalibration.

Homogeneity.

Drift.

Thermoelement diameter, roundness, and surface appearance.

Thermoelement spacing.

Thermoelement ductility.

Metallurgical structure.

1.3.4 Thermocouple Assembly PropertiesDimensions-length, diameter, and roundness.

Surface-gross visual, finish, reference end seal, and defect detection by dye penetrant.

Electrical-continuity, loop resistance, and connector polarity.

Radiographic inspection.

Thermoelement diameter.

Thermal response time.

Thermal cycle.

Ключевые слова:
high temperature insulation test; insulation resistance; sheath integrity; sheathed thermocouple; thermal cycle; thermal response time; thermocouple; thermocouple assembly properties; thermoelement; thermoelement properties