Formerly under the jurisdiction of Committee D19 on Water, these test methods were withdrawn in July 2020 in accordance with section 10.6.3 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.
Значение и использованиеThese test methods provide information on the dimensional stability and wetting characteristics of autoclaved membrane filters. Excessive dimensional changes caused by autoclaving a constrained membrane filter could crack the filter, rendering it useless. Increased hydrophobicity, due to autoclaving would result in an increased time to filter a liquid sample.
Область применения1.1 These test methods deal with tests to determine the autoclavability of membrane filters.
1.2 These test methods are applicable to all membrane filters. Either Test Method A, Test Method B, or both, may be utilized to assess the autoclavability of membrane filters.