5.1 This test method was developed to measure the concentration of 210Pb in nonprocess water samples. This test method may be used to determine the concentration of 210Pb in environmental samples.
Область применения1.1 This test method covers the determination of radioactive 210Pb in environmental water samples (for example, drinking, non-process and effluent waters) in the range of 37 mBq/L (1.0 pCi/L), but that can be adjusted lower or higher (by varying sample volume or count duration, or both) as needed to meet measurement quality objectives.
1.2 The values stated in SI units are to be regarded as standard. The values given in parentheses are provided for information purposes only.
1.3 This method has been used successfully with tap water. It is the user's responsibility to ensure the validity of this test method for samples larger than 500 mL and for waters of untested matrices.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. For specific hazards statements, see Section 9.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.