Каталог стандартов
+7 (495) 223-46-76
+7 (812) 309-78-59
inform@normdocs.ru
1.1 This test method covers the techniques of thermocouple calibration based upon comparisons of thermocouple indications with those of a reference thermometer, different from methods involving the use of fixed points. The precise evaluation of the electromotive force (emf)-temperature relation of a thermocouple is accomplished by determining its emf output at each of a series of measured temperatures. Calibrations are covered over temperature ranges appropriate to the individual types of thermocouples within an over-all range from about -180 to 1700°C (-290 to 2660°F).
1.2 In general, the test method is applicable to bare wire thermocouples or sheathed thermocouples. The latter may require special care to control thermal conduction losses.