ASTM E2254-08

Заменен
Standard Test Method for Storage Modulus Calibration of Dynamic Mechanical Analyzers — 3 стр.
Значение и использование

This test method calibrates or demonstrates conformity of a dynamic mechanical analyzer at an isothermal temperature within the range of -100 to 300 °C.

Dynamic mechanical analysis experiments often use temperature ramps. This method does not address the effect of that change in temperature on the storage modulus.

A calibration factor may be required to obtain corrected storage modulus values.

This method may be used in research and development, specification acceptance, and quality control or assurance.

Область применения

1.1 This test method describes the calibration or performance confirmation for the storage modulus scale of a commercial or custom built dynamic mechanical analyzer (DMA) over the temperature range of -100 to 300 °C using reference materials in the range of 1 to 200 GPa.

1.2 SI units are the standard.

1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ICS
17.020 Metrology. Including measuring instruments in general, preferred numbers, standard measures, general aspects of reference materials, etc. / Метрология и измерения в целом. Включая измерительные приборы в целом, предпочтительные числа, эталонные меры, общие аспекты эталонных материалов и т.д.
Сборник ASTM
14.05 Statistical Methods; Hazard Potential of Chemicals; Thermal Measurements; Manufacture of Pharmaceutical and Biopharmaceutical Products / Статистические методы; Потенциальная опасность химических веществ; Измерения теплопроводности; Производство фармацевтических изделий
Тематика
Thermal Analysis