This specification covers the general workmanship aspects of inserted and noninserted 410, 410K, 416, 420, 420A, and 420B martensitic stainless steel scissors intended for reuse in surgery. Inserts should be made of stellite, tungsten carbide, or other suitable materials. The instruments should conform to the required values of corrosion resistance cutting ability, Rockwell hardness, and physical properties after passivation and heat treatment. The products should also have finger rings, joints, cutting edges, pivot screws, and finish that are of the required quality.
Область применения1.1 This specification covers general workmanship aspects of inserted and noninserted stainless steel scissors fabricated from stainless steel and intended for reuse in surgery.
1.2 The following safety hazards caveat pertains only to the test method described in this specification: This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.