Каталог стандартов

+7 (495) 223-46-76 +7 (812) 309-78-59
inform@normdocs.ru

ASTM F1208-89(2005)

Отменен
Standard Specification for Minimum Performance and Safety Requirements for Anesthesia Breathing Systems (Withdrawn 2014) — 13 стр.
Причина отмены

This specification covers breathing systems and components employed with anesthesia gas machines for humans. This specification considers the circle system as a whole and some of its components individually. A particular emphasis is placed upon component arrangement in the circle absorber-type system, and submits a system of standard description and notation. Excluded are ventilators for use during anesthesia, Mapelson nonrebreathing type systems, as well as breathing systems and related components of dental analgesia machines.

Formerly under the jurisdiction of Committee F29 on Anesthetic and Respiratory Equipment, this specification was withdrawn in July 2014 in accordance with section 10.6.3 of the Regulations Governing ASTM Technical Committees, which requires that standards shall be updated by the end of the eighth year since the last approval date.

Реферат

This specification covers breathing systems and components employed with anesthesia gas machines for humans. This specification considers the circle system as a whole and some of its components individually. A particular emphasis is placed upon component arrangement in the circle absorber-type system, and submits a system of standard description and notation. The circle system is divided into four segments: segment 1 - from patient to expiratory valve, segment 2 - from expiratory valve to reservoir bag, segment 3 - from reservoir bag to inspiratory valve, and segment 4 - from inspiratory valve to patient. Leakage of breathing system and resistance of breathing system shall be tested to meet the requirements prescribed. The adjustable pressure-limiting valves, reservoir bag connectors and reservoir bags, carbon dioxide absorbers, unidirectional valves, and optional components shall meet the requirements prescribed.

Область применения

1.1 This specification covers breathing systems and components employed with anesthesia gas machines for humans. This specification considers the circle system as a whole and some of its components individually. A particular emphasis is placed upon component arrangement in the circle absorber-type system, and submits a system of standard description and notation. Excluded are ventilators for use during anesthesia, Mapelson nonrebreathing type systems, as well as breathing systems and related components of dental analgesia machines. (For rationale, see Appendix X1.)

1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

1.3 This specification is arranged as follows:

SectionScopeReferenced DocumentsTerminologySystem ClassificationGraphic NotationsDiagramatic RulesFunctional Segments of the Circle SystemTest Procedures-GeneralSterilizationSystemsLeakage of Breathing SystemResistance of Breathing SystemsVolume of Gas Not Delivered to Patient Due to Internal ComplianceMonitoring RequirementsAdjustable Pressure-Limiting (APL) ValvesReservoir BagsCO2 AbsorbersUnidirectional ValvesRequirementsTest ProcedureOptional ComponentsBreathing TubesFresh Gas Inlet and Fresh Gas Supply Tube or HoseConnectorsPressure GaugesY-Piece and Breathing TubesInformation in LabelingRationale

ICS
11.040.10 Anaesthetic,. Including medical gas installations / Наркозные, дыхательные и реанимационные аппараты
Сборник ASTM
13.02 Medical and Surgical Materials and Devices (II): F2502–Latest; Emergency Medical Services; Search and Rescue; Anesthetic and Respiratory Equipment / Медицинские и хирургические материалы и приборы (II): F2502-Latest; Скорая медицинская помощь; Поиск и Спасение; Оборудование для анестезии и искусственного дыхания
Тематика
Medical Services