Каталог стандартов

+7 (495) 223-46-76 +7 (812) 309-78-59
inform@normdocs.ru

ASTM F1472-14

Заменен
Standard Specification for Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical Implant Applications (UNS R56400) — 5 стр.
Реферат

This specification covers the requirements for wrought titanium-6aluminum-4vanadium alloy for use in surgical implant applications. The material shall be manufactured by multiple melting in plasma arc and electron beam furnaces into strip, sheet, plate, bar, forging bar, or wire. The material shall be available in the annealed or cold-worked condition, with the surface mechanically descaled or pickled, sandblasted, chemically milled, ground, machined, peeled, polished, or a combination of these. Heat and product analyses as well as tension and bend (for sheet, strip, and plate) tests shall be performed and shall conform to the requirements specified. In addition, the material shall have a microstructure consisting of a fine dispersion of alpha and beta phases as a result of processing in the alpha plus beta field, with no continuous alpha network at prior beta grain boundaries and no coarse, elongated alpha platelets.

Область применения

1.1 This specification covers the chemical, mechanical, and metallurgical requirements for wrought annealed titanium-6aluminum-4vanadium alloy (UNS R56400) to be used in the manufacture of surgical implants.

1.2 Units—The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.

ICS
11.040.40 Implants. Including pacemakers / Имплантаты для хирургии, протезирования и ортоптики. Включая кардиостимуляторы
Сборник ASTM
13.01 Medical and Surgical Materials and Devices (I): E667 – F2477 / Медицинские и хирургические материалы и приборы (I): E667 - F2477
Тематика
Medical Devices & Implants