Каталог стандартов

+7 (495) 223-46-76 +7 (812) 309-78-59
inform@normdocs.ru

ASTM F1830-97(2005)

Заменен
Standard Practice for Selection of Blood for <i>in vitro</i> Evaluation of Blood Pumps — 2 стр.
Значение и использование

The test results are substantially affected by donor species and age, the method of harvesting, the period of storage, the biochemical state of the blood, and the hemoglobin and hematocrit level of blood. , Therefore, standardization of proper blood usage for in vitro evaluation of blood pumps is essential, and this recommended practice will allow a universal comparison of test results.

Drawing several units of blood from healthy cattle does not affect them or their health. Therefore, bovine blood is strongly suggested for usage in experimental evaluation of blood damage. Mixing two donor sources of blood should be avoided in hemolysis tests because the mixture may induce added hemolysis or a change in red cell resistance against trauma.

Область применения

1.1 This practice covers blood that will be used for in vitro performance assessments of blood pumps. These assessments include the hemolytic properties of the devices.

1.2 This practice covers the utilization of blood for the in vitro evaluation of the following devices:

1.2.1 Continuous flow rotary blood pumps (roller pumps, centrifugal pumps, axial flow pumps, and so forth) (see Practice F1841).

1.2.2 Pulsatile blood pumps (pneumatically driven, electromechanically driven, and so forth).

1.3 The source of blood utilized for in vitro evaluation of blood trauma (that is, hemolysis caused by the blood pumps, due to the pump design, construction, and materials used) substantially influences the results of the performance of these devices. Thus, a standardized blood source is required.

ICS
11.100 Laboratory medicine / Лабораторная медицина
Сборник ASTM
13.01 Medical and Surgical Materials and Devices (I): E667 – F2477 / Медицинские и хирургические материалы и приборы (I): E667 - F2477
Тематика
Medical Devices & Implants