Каталог стандартов

+7 (495) 223-46-76 +7 (812) 309-78-59
inform@normdocs.ru

ASTM F2096-02e1

Заменен
Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test) — 5 стр.
Область применения

1.1 This test method covers the detection of gross leaks in medical packaging. Method sensitivity is down to 250 μm with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.

1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.

1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure

1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.

1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

ICS
55.040 Packaging. Including wrappers, paper, films, foils, cords, sealing devices, cushioning, etc. / Упаковочные материалы и приспособления. Включая обертки, бумагу, пленки, фольгу, веревки, устройства для запечатывания, прокладки и т.д.
19.060 Mechanical. Including testing equipment / Механические испытания. Включая испытательное оборудование
Сборник ASTM
15.10 Packaging; Primary Barrier Packaging; Cannabis / Упаковка; Первичная барьерная упаковка; Каннабис
Тематика
Paper & Packaging