ASTM F749-98(2012)

Заменен
Standard Practice for Evaluating Material Extracts by Intracutaneous Injection in the Rabbit — 3 стр.
Значение и использование

4.1 This practice is to be used to help assess the biocompatibility of materials used in medical devices. It is an acute toxicological test designed to detect the presence of injurious leachable substances.

4.2 This practice may not be appropriate for all types of implant applications. The user is cautioned to consider the appropriateness of the method in view of the materials being tested, their potential applications, and the recommendations contained in Practice F748.

4.3 The only applicable limitation is the extract preparation. Refer to Sections 4.3 and 4.4 of Practice F619 for a description of this limitation.

Область применения

1.1 This practice is a nonspecific, acute toxicity test used to help determine the biocompatibility of materials used in medical devices.

1.2 The liquids injected into the rabbits are those obtained by Practice F619 where the extraction vehicles are saline, vegetable oil, or other liquids simulating human body fluids.

1.3 This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 may provide guidance for the selection of appropriate methods for testing materials for a specific application.

1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

ICS
11.220 Veterinary. Including equipment specific to veterinary medicine / Ветеринария. Включая специальное ветеринарное оборудование
Сборник ASTM
13.01 Medical and Surgical Materials and Devices (I): E667 – F2477 / Медицинские и хирургические материалы и приборы (I): E667 - F2477
Тематика
Medical Devices & Implants