This specification covers the chemical, mechanical, and metallurgical requirements for wrought cobalt-20chromium-15tungsten-10nickel alloy bars, rods, wires, sheets, and strips (except surgical fixation wires) for use in surgical implants. All alloys shall be furnished to the purchaser's specifications, in the annealed or cold-worked condition. Then, bars and wires shall be finished bright annealed, cold drawn, pickled, ground, or ground and polished, as specified by the purchaser. While, sheets shall be furnished bright annealed, pickled, cold-rolled, or polished, as specified by the purchaser. The alloys shall adhere to specified values of ultimate tensile strength, yield strength, elongation, and microcleanliness.
Область применения1.1 This specification covers the requirements for wrought cobalt-20chromium-15tungsten-10nickel alloy used for surgical implants. The properties specified apply specifically to wrought bar, rod, wire, sheet, and strip, but do not apply to surgical fixation wire (see Specification F1091).
1.2 The SI units in this standard are the primary units. Values in primary SI or secondary inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore each system shall be used independent of the other. Combining values from the two systems may result in non-conformance with the specification.