Каталог стандартов

+7 (495) 223-46-76 +7 (812) 309-78-59
inform@normdocs.ru

ÖNORM EN 26777:1993-05

Действует
Water quality - Determination of nitrite - Molecular absorption spectrometric method (ISO 6777:1984)
2.1 Range
A nitrite nitrogen concentration, rho N, of up to 0.25 mg/l can be
determined when using the maximum volume (40 ml) of test
portion.
2.2 Limit of detection
When using cells of optical path length 40 mm and a test portion
of 40 ml, the limit of detection has been determined to lie
within the range rho N = 0,001 to 0,002 mg/l.
2.3 Sensitivity
Using a 40 ml test portion and a cell of optical path length
40 mm, rho N = 0,062 mg/l gives an absorbance of about
0,66 units.
Using a 40 ml test portion and a cell of optical path length
10 mm, rho N = 0,25 mg/l gives an absorbance of about
0,67 units.
2.1 Arbeitsbereich
Wird das höchstzulässige Analysenproben-Volumen (40 ml)
eingesetzt, kann eine Nitritstickstoff-Konzentration, rho N, bis zu
0,25 mg/l bestimmt werden.
2.2 Nachweisgrenze
Bei Verwendung von Küvetten der optischen Weglänge
von 40 mm und einem Analysenproben-Volumen von 40 ml
wurde eine Nachweisgrenze zwischen rho N = 0,001 und
0,002 mg/l bestimmt.
2.3 Empfindlichkeit
Bei einem Analysenproben-Volumen von 40 ml und Einsatz
von Küvetten der optischen Weglänge von 40 mm ergab
rho N = 0,062 mg/l eine Extinktion von etwa 0,66 Einheiten.
Bei einem Analysenproben-Volumen von 40 ml und Einsatz
von Küvetten der optischen Weglänge von 10 mm ergab
rho N = 0,25 mg/l eine Extinktion von etwa 0,67 Einheiten.
ICS