This European Standard specifies procedures for the evaluation of basic performance parameters of the radiographic film digitisation process such as spatial resolution and spatial linearity, density range, density contrast sensitivity and characteristic transfer curve. They can be integrated into the system software and together with a standard reference film (as described in clause 5) used for quality control of the digitisation process. This reference film provides a series of test targets for performance evaluation. The test targets are suitable for evaluating a digitisation system with a spatial resolution down to 25 µm, a density contrast sensitivity down to 0,02 optical density, a density range of 0,5 to 4,5 and a film size capacity of (350 x 430) mm². This standard does not address signal processing and display of the digitised data.
Diese Europäische Norm legt Messvorschriften für die Bestimmung von Basisparametern für Röntgenfilm-Digitalisierungssysteme wie Ortsauflösung und geometrische Linearität, Umfang der optischen Dichte, Dichtekontrastempfindlichkeit und charakteristische Übertragungskurve fest. Diese Messvorschriften können in die Systemsoftware integriert und in Verbindung mit einem Standard-Referenzfilm (wie in Abschnitt 5 beschrieben) zur Qualitätssicherung des Digitalisierungsvorganges benutzt werden. Dieser Referenzfilm verfügt über eine Reihe von Testmustern zur Bestimmung der Leistungsfähigkeit des Digitalisierungssystems. Die Testmuster sind geeignet für die Überprüfung von Digitalisierungssystemen mit einer Ortsauflösung von maximal 25 µm, einer Dichtekontrastempfindlichkeit von 0,02 optische Dichten, einem Bereich der optischen Dichte (Filmschwärzung) zwischen 0,5 und 4,5 und einer maximalen Filmgröße von (350 × 430) mm². Diese Norm behandelt weder die Signalverarbeitung noch die Darstellung der digitalisierten Daten.