This International Standard specifies a method of dynamic testing of single post endosseous dental implants of the transmucosal type in combination with their premanufactured prosthetic components. It is most useful for comparing endosseous dental implants of different designs or sizes. This International Standard is not a test of the fundamental fatigue properties of the materials from which the endosseous implants and prosthetic components are made.
This International Standard is not applicable to dental implants with endosseous lengths shorter than 8 mm nor to magnetic attachments.
While this International Standard simulates the functional loading of an endosseous dental implant under "worst case" conditions, it is not applicable for predicting the in vivo performance of an endosseous dental implant or dental prosthesis, particularly if multiple endosseous dental implants are used for a dental prosthesis.
Diese Internationale Norm legt ein Prüfverfahren für die dynamische Prüfung von einpfostigen enossalen Dentalimplantaten vom transmukosalen Typ zusammen mit ihren vorgefertigten Prothetikteilen fest. Das Verfahren ist zum Vergleich von enossalen Dentalimplantaten mit verschiedenen Konstruktionen oder Größen am besten geeignet. Diese Internationale Norm ist keine Prüfung, um die grundlegenden Ermüdungseigenschaften der Materialien zu ermitteln, aus denen Implantat und Prothetikteile gefertigt sind.
Diese Internationale Norm ist nicht anwendbar auf Dentalimplantate mit enossalen Längen kürzer als 8 mm oder auf magnetische Retentionselemente.
Obwohl in dieser Internationalen Norm die funktionelle Belastung eines enossalen Dentalimplantates unter den Bedingungen des "ungünstigsten Falls" (en: worst case) simuliert wird, ist diese Norm nicht anwendbar zur Voraussage der in vivo Leistungsfähigkeit eines enossalen Dentalimplantats oder eines Zahnersatzes, besonders wenn für einen Zahnersatz mehrere enossale Dentalimplantate verwendet werden.