This International Standard specifies requirements for torque transmitters to be used in oral implantology in conjunction with a dental handpiece as an accessory in the placement of dental implants and the further manipulation of connecting parts in the craniofacial area.
This International Standard applies to torque transmitters used for placement and for removal in the oral cavity of the patient which are connected to power-driven systems having torque control mechanism, but does not apply to the power-driven systems themselves.
This International Standard does not include the dental implant nor parts that would be connected to it.
With regard to safety, this International Standard gives requirements for classification, intended performance, performance attributes, material selection, performance evaluation, manufacture, reprocessing and information to be supplied by the manufacturer.
Diese Internationale Norm legt Anforderungen an Drehmomentübertragungen fest, die in der oralen Implantologie in Verbindung mit einem zahnärztlichen Handstück als Zubehör beim Einsetzen von Dentalimplantaten und der weiteren Handhabung von Verbindungsteilen im Schädel-Gesichts-Bereich verwendet werden.
Diese Internationale Norm gilt für Drehmomentübertragungen, die zum Einsetzen und Entfernen in der Mundhöhle des Patienten verwendet werden und die an elektrische Antriebssysteme mit einer Drehmomentsteuerung angeschlossen sind, aber sie gilt nicht für die elektrischen Antriebssysteme selbst.
Diese Internationale Norm enthält nicht das Dentalimplantat sowie Teile, die mit diesem verbunden werden können.
Für die Sicherheit legt diese Internationale Norm Anforderungen an Einteilung, vorgesehene Leistung, Leistungsmerkmale, Materialauswahl, Leistungsbewertung, Herstellung, Wiederaufbereitung und die vom Hersteller bereitzustellenden Angaben fest.