This document specifies a standard reference zero for the scale of hearing threshold level applicable to pure-tone air conduction audiometers, to promote agreement and uniformity in the expression of hearing threshold level measurements throughout the world.
It states the information in a form suitable for direct application to the calibration of audiometers, that is, in terms of the reference equivalent threshold sound pressure levels of generic supra-aural earphones specified in 4.2, measured on an ear simulator complying with IEC 60318-1 and in terms of model-specific data given in two additional tables for the IEC 60318-3 acoustic coupler and the IEC 60318-1 ear simulator, respectively.
The data are based on an assessment of the information available from the various standardizing laboratories responsible for audiometric standards and from scientific publications.
Dieses Dokument legt Standard-Bezugspegel für die Hörschwellenpegel-Skala von Reinton-Luftleitungs-Audiometern fest, um die weltweite Übereinstimmung und Einheitlichkeit in der Angabe der Ergebnisse von Hörschwellenpegel-Messungen zu fördern.
Es gibt die Daten in einer solchen Weise an, dass sie bei der Kalibrierung von Audiometern unmittelbar angewendet werden können, d. h. in Form äquivalenter Bezugs-Schwellenschalldruckpegel von in 4.2 festgelegten typischen supra-auralen Kopfhörern, die an einem Ohrsimulator nach IEC 60318-1 gemessen werden, und in Form bauartspezifischer Daten, welche in zwei zusätzlichen Tabellen für akustische Kuppler nach IEC 60318-3 bzw. für Ohrsimulatoren nach IEC 60318-1 angegeben sind.
Die Daten beruhen auf der Auswertung von Angaben verschiedener für die audiometrischen Normale verantwortlicher Laboratorien und wissenschaftlicher Veröffentlichungen.