This part of EN 45502 specifies requirements that are generally applicable to ACTIVE IMPLANTABLE MEDICAL
DEVICES.
NOTE 1 For particular types of ACTIVE IMPLANTABLE MEDICAL DEVICES, these general requirements are supplemented or
modified by the requirements of particular standards which form additional parts of this European Standard.
The tests that are specified in EN 45502 are type tests and are to be carried out on samples of an ACTIVE
IMPLANTABLE MEDICAL DEVICE to show compliance.
This part of EN 45502 is applicable not only to ACTIVE IMPLANTABLE MEDICAL DEVICES that are electrically
powered but also to those powered by other energy sources (for example by gas pressure or by springs).
This part of EN 45502 is also applicable to some non-implantable parts and accessories of the ACTIVE
IMPLANTABLE MEDICAL DEVICES.
NOTE 2 The device that is commonly referred to as an ACTIVE IMPLANTABLE MEDICAL DEVICE can be a single device, a
combination of devices, or a combination of a device or devices and one or more accessories. Not all of these parts are
required to be either partially or totally implantable, but there is a need to specify some requirements of non-implantable
parts and accessories if they could affect the safety or performance of the implantable device.
NOTE 3 In this part of EN 45502, terms printed in small capital letters are used as defined in Clause 3. Where a defined
term is used as a qualifier in another term, it is not printed in small capital letters unless the concept thus qualified is also
defined.